WATER BATHS
SWB30
SWB30-2
Previously Designated as
WPC95 & WPC95-2
MICROPROCESSOR CONTROLLED
INSTALLATION AND OPERATION MANUAL
4861564
Revised June 19, 2014
Zobrazit stránku 0
1 2 3 4 5 6 ... 17 18

Shrnutí obsahu

Strany 1 - WATER BATHS

WATER BATHS SWB30 SWB30-2 Previously Designated as WPC95 & WPC95-2 MICROPROCESSOR CONTROLLED INSTALLATION AND OPERATION MANU

Strany 2

6 10 Microprocessor Control Water Baths chamber temperature has been stable at the set point for several hours. 1. Place a calibrated referen

Strany 3 - TABLE OF CONTENTS

6 Microprocessor Control Water Baths 11 Bath Cover Assembly To assemble the bath cover, perform the following steps: 1. Insert 2 (Item 4) self-tapp

Strany 4 - INTRODUCTION

Microprocessor Control Water Baths 12 MAINTENANCE Warning: Prior to any maintenance or service on this unit, disconnect the power cord from the

Strany 5 - RECEIVING YOUR UNIT

13 Microprocessor Control Water Baths TROUBLESHOOTING Should the unit malfunction, use this section to determine the problem and resolution. Troub

Strany 6 - GRAPHIC SYMBOLS

8 14 Microprocessor Control Water Baths Problem Possible Cause Solution Unit will not heat up at all Controller Fault. Do all controller functions

Strany 7 - INSTALLATION

15 Microprocessor Control Water Baths PARTS LIST Table 3. Parts Description 100-120V 220-240V Element 9570743 9570744 Fuse, 6.3 AMP 3300515 33005

Strany 8 - CONTROLS OVERVIEW

16 Microprocessor Control Water Baths UNIT SPECIFICATIONS Table 4. Weight Table 5. Dimensions Table 6. Capacity Table 7. Temperature Tab

Strany 9 - OPERATION

Microprocessor Control Water Baths 17 SCHEMATICS Figure 4. Wire Diagram SWB30 (WPC95) (9851341) HOTNEUT RALGROUNDLOA DSENSOR124TEMPERATURECONTRO

Strany 10 - 9850563

18 Microprocessor Control Water Baths Figure 6. Wire Diagram SWB30-2 (WPC95-2) (9851342) HOTNEUTRALGROUNDLOADSENSOR124TEMPERATURECONTROL1750966O

Strany 11 - Bath Cover Assembly

2 Microprocessor Control Water Baths These units are TUV CUE listed as water baths for professional, industrial or educational use where the preparat

Strany 12 - MAINTENANCE

Microprocessor Control Water Baths 3 TABLE OF CONTENTS INTRODUCTION ...

Strany 13 - TROUBLESHOOTING

Microprocessor Control Water Baths 4 INTRODUCTION Thank you for choosing a general purpose water bath. These units are not intended for use at haz

Strany 14 - Service

5 Microprocessor Control Water Baths RECEIVING YOUR UNIT Before leaving our factory, all units are packaged in high quality shipping materials des

Strany 15 - PARTS LIST

Microprocessor Control Water Baths 6 GRAPHIC SYMBOLS Your unit is provided with a display of graphic symbols that should help in identifying user

Strany 16 - UNIT SPECIFICATIONS

7 Microprocessor Control Water Baths INSTALLATION Your satisfaction and safety require a complete understanding of this unit. Read the instructi

Strany 17 - SCHEMATICS

Microprocessor Control Water Baths 8 CONTROLS OVERVIEW This section provides an overview of the panel controls. See Figure 1 for an illustration

Strany 18 - NEUTRAL TERMINAL BLOCK

Microprocessor Control Water Baths 9 OPERATION Warning: These baths are not intended for use as acid baths. Use as an acid bath will cause sever